𝔖 Bobbio Scriptorium
✦   LIBER   ✦

NH3/O2 mixed gas plasmas alter the interaction of blood components with stainless steel

✍ Scribed by Chen, Meng ;Zamora, Paul O. ;Peña, Louis ;Som, Prantika ;Osaki, Shigemasa


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
173 KB
Volume
67A
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Stainless steel treated with a mixed gas plasma of NH~3~ plus O~2~ had chemical and biologic characteristics distinct from untreated stainless steel or stainless steel treated with NH~3~ or O~2~ plasmas used separately. NH~3~/O~2~ plasmas deposited nitrogen as both −CN (organic) and −NO (nitrate, nitrite)—materials not found on untreated stainless steel—and the contact angle changed from 44° to 23°. Treatment of stainless steel (and titanium) resulted in surfaces with enhanced resistance to platelet and leukocyte attachment. A gas plasma of N~2~O/O~2~ also was found to reduce platelet and leukocyte attachment, suggesting that these properties may be common to surfaces coated with oxynitrites (nitrides). Upon subcutaneous implantation, no inflammation, hemolysis, or untoward thrombosis was noted in the tissue surrounding the wafers treated with the NH~3~/O~2~ plasmas, although the cellular density was considerably reduced by 2 weeks after implant. Collectively, the results suggest that NH~3~/O~2~ plasmas impart a unique character to stainless steel that may be useful in the construction of medical devices. © 2003 Wiley Periodicals, Inc. J Biomed Mater Res 67A: 994–1000, 2003