𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Myocardial T1 mapping with MRI: Comparison of look-locker and MOLLI sequences

✍ Scribed by Marcelo Souto Nacif; Evrim B. Turkbey; Neville Gai; Saman Nazarian; Rob J. van der Geest; Radwa A. Noureldin; Christopher T. Sibley; Martin Ugander; Songtao Liu; Andrew E. Arai; João A.C. Lima; David A. Bluemke


Book ID
102904975
Publisher
John Wiley and Sons
Year
2011
Tongue
English
Weight
505 KB
Volume
34
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose:

To evaluate the relationship between “Look‐Locker” (LL) and modified Look‐Locker Inversion recovery (MOLLI) approaches for T1 mapping of the myocardium.

Materials and Methods:

A total of 168 myocardial T1 maps using MOLLI and 165 maps using LL were obtained in human subjects at 1.5 Tesla. The T1 values of the myocardium were calculated before and at five time points after gadolinium administration. All time and heart rate normalizations were done. The T1 values obtained were compared to determine the absolute and bias agreement.

Results:

The precontrast global T1 values were similar when measured by the LL and by MOLLI technique (mean, 1004.9 ms ± 120.3 versus 1034.1 ms ± 53.1, respectively, P = 0.26). Postcontrast myocardial T1 time from LL was significantly longer than MOLLI from 5 to 25 min (mean difference, LL ‐ MOLLI was +61.8 ± 46.4 ms, P < 0.001). No significant differences in T1 values were noted between long and short axis measurements for either MOLLI or LL.

Conclusion:

Postcontrast LL and MOLLI showed very good agreement, although LL values are higher than MOLLI. Precontrast T1 values showed good agreement, however LL has greater limits of agreement. Short and long axis planes can reliably assess T1 values. J. Magn. Reson. Imaging 2011;. © 2011 Wiley Periodicals, Inc.


📜 SIMILAR VOLUMES


Modified Look-Locker inversion recovery
✍ Daniel R. Messroghli; Aleksandra Radjenovic; Sebastian Kozerke; David M. Higgins 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 788 KB

A novel pulse sequence scheme is presented that allows the measurement and mapping of myocardial T1 in vivo on a 1.5 Tesla MR system within a single breath-hold. Two major modifications of conventional Look-Locker (LL) imaging are introduced: 1) selective data acquisition, and 2) merging of data fro