✦ Synopsis


Background

Late Gadolinium Enhancement (LGE) and T2-weighted cardiovascular magnetic resonance (CMR) provides a means to measure myocardial area at risk (AAR) and salvage. Several T2-weighted CMR sequences are in use, but there is no consensus in terms of which sequence to be the preferred. Therefore, the aim of the present study was to: (1) Assess the reproducibility and (2) compare the two most frequently used T2-weighted CMR protocols for measuring AAR and salvage.

Methods

91 patients with ST-elevation myocardial infarction treated with primary percutaneous coronary intervention underwent a CMR scan 1-7 days after initial treatment. Two different T2-weighted protocols, varying in slice thickness and echo time (TE), were applied covering the entire left ventricle (LV) (protocol 1: TE 65 msec and slice thickness 15 mm; protocol 2: TE 100 msec and slice thickness of 8 mm). On a second scan performed 3 months later, infarct size was assessed with a standard LGE sequence. The two protocols were compared in terms of AAR and salvage index. Furthermore, intra- and interobserver reproducibility were assessed.

Results

Protocol 1 measures a larger AAR and salvage index than protocol 2 with a mean difference in AAR of 1 ± 8%LV (p < 0.01) and 6 ± 12 g (p < 0.01) and salvage index of 0.04 ± 0.12 (p < 0.01). Both protocols had a high intra- and interobserver reproducibility with acceptable limits of agreement (6-8%LV and 6-12 g in AAR and 0.06-0.08 in salvage index).

Conclusions

We report acceptable reproducibility for AAR and salvage index measured by T2-weighted images. Thus CMR is a reliable tool for measuring AAR and salvage index. Protocol 2 (8 mm slice thickness and 100 msec TE) measures slightly smaller AAR than protocol 1 (15 mm slice thickness and 65 msec TE), but the present study does not allow for a clear recommendation of either of the protocols.


📜 SIMILAR VOLUMES


Myocardium at risk in ST-elevation myoca
✍ Georg F Fuernau; Ingo Eitel; Vinzenz Franke; Lysann Hildebrandt; Josefine Meissn 📂 Article 📅 2011 🏛 BioMed Central 🌐 English ⚖ 142 KB

## Objectives T2-weighted magnetic resonance imaging (MRI) and the endocardial surface area (ESA) assessed by late gadolinium enhanced (LGE) have been introduced as relatively new methods for myocardium at risk assessment in ST-elevation myocardial infarction (STEMI).

Myocardium at risk by magnetic resonance
✍ Joey F Ubachs; Peder Sorensson; Henrik Engblom; Marcus Carlsson; Stefan Jovinge; 📂 Article 📅 2012 🏛 BioMed Central 🌐 English ⚖ 83 KB

Ubachs et al.: Myocardium at risk by magnetic resonance imaging: head-to-head comparison of T2-weighted imaging and early gadolinium enhanced steady state free precession.