## Abstract Noise properties, the signal‐to‐noise ratio (SNR), contrast‐to‐noise ratio (CNR), and signal responses were compared during functional activation of the human brain at 1.5 and 3.0 T. At the higher field spiral gradient‐echo (GRE) brain images revealed an average gain in SNR of 1.7 in fu
Myelin water imaging: Implementation and development at 3.0T and comparison to 1.5T measurements
✍ Scribed by Shannon H. Kolind; Burkhard Mädler; Stefan Fischer; David K.B. Li; Alex L. MacKay
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 727 KB
- Volume
- 62
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Multicomponent T~2~ relaxation imaging can be used to measure signal from water trapped between myelin bilayers; the ratio of myelin water signal to total water is termed the myelin water fraction (MWF). The goal of this study was to implement and develop the single‐slice T~2~‐imaging technique proposed by Poon and Henkelman. For refinement, scan parameters (gradient crusher height and slew rate, bandwidth, echo spacing, matrix size, repetition time, and phase rewinding) were varied in water‐based phantoms and in fixed and in vivo brain. Changes in the standard deviation of the residuals of the multiexponential fit, MWF, T~2~, and peak width of the intra/extracellular water were monitored to determine which scan parameters minimized artifacts. Subsequently, we compared multicomponent T~2~ measurements at 1.5T and 3.0T for 10 healthy volunteers, and investigated the differences in SNR, fit residuals, MWF, and T~2~ and peak width of the intra/extracellular water, at higher magnetic field. MWF maps were found to be qualitatively similar between field strengths. MWFs were found to be significantly higher at 3.0T than at 1.5T, but with a strongly significant correlation between measurements (R^2^ > 0.92, P < 0.0005). The signal‐to‐noise ratio (SNR) was nearly double at 3.0T, but the standard deviation of residuals was increased in most cases. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Assessment of lung effective transverse relaxation time (__T__~2~\*) may play an important role in the detection of structural and functional changes caused by lung diseases such as emphysema and chronic bronchitis. While __T__~2~\* measurements have been conducted in both animals and h
## Abstract ## Purpose To compare two multislice turbo spin‐echo (TSE) carotid artery wall imaging techniques at 1.5 T and 3.0 T, and to investigate the feasibility of higher spatial resolution carotid artery wall imaging at 3.0 T. ## Materials and Methods Multislice proton density‐weighted (PDW
## Abstract ## Purpose To determine normative R2\* values in the liver and heart at 3T, and establish the relationship between R2\* at 3T and 1.5T over a range of tissue iron concentrations. ## Materials and Methods A total of 20 healthy control subjects and 14 transfusion‐dependent patients wer
## Abstract ## Purpose To quantitatively compare spiral imaging and sensitivity‐encoded‐echo‐planar‐imaging (SENSE‐EPI) methods for blood oxygen level‐dependent (BOLD) imaging using controlled changes in the end‐tidal partial pressure of CO~2~ (PetCO~2~) to provide a global BOLD response. Specific
## Abstract Most macromolecular contrast agents (CAs) show lower r~1~ and higher r~2~ relaxivities at 3.0T than at 1.5T. MR lymphangiography in mice using a macromolecular G6 dendrimer‐based CA was serially performed and compared at both 1.5T and 3.0T. The r~1~ and r~2~ relaxivities of the G6 CA we