𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Multisyringe flow-injection system for total inorganic arsenic determination by hydride generation-atomic fluorescence spectrometry

✍ Scribed by N.V Semenova; L.O Leal; R Forteza; V Cerdà


Publisher
Elsevier Science
Year
2002
Tongue
English
Weight
161 KB
Volume
455
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


A software-controlled time-based multisyringe flow-injection system for total inorganic arsenic determination by hydride generation atomic fluorescence spectrometry (HGAFS) has been developed. By using a multisyringe burette coupled with one multiport selection valve, the time-based injection provides precise known volumes of sample, a reducing sodium tetrahydroborate solution and a pre-reducing solution which are dispensed into a gas-liquid separation cell. An argon flow delivers the arsine into the flame of an atomic fluorescence spectrometer. A hydrogen flow has been used to support the flame.

Linear calibration graphs for arsenic concentrations between 0.25 and 12 g l -1 were obtained. The detection limit of the proposed technique (3σ b /S) was 0.07 g l -1 . A sample throughput of 36 samples/h (108 injections) has been achieved. The proposed technique has been validated by means of reference solid and water materials with good agreement with the certified values. This method was compared with those reported in previous sequential injection analysis (SIA) and flow-injection analysis (FIA) systems. The proposed method offers a number of advantages in front the usual AFS applications, which are mainly a higher sampling frequency and a significant reduction in reagent consumption.


📜 SIMILAR VOLUMES


A flow injection/hydride generation syst
✍ R.R. Liversage; J.C. Van Loon; J.C. De Andrade 📂 Article 📅 1984 🏛 Elsevier Science 🌐 English ⚖ 587 KB

A flow-injection system is described for the semi-automatic determination of arsenic (10-1000 pg 1-l) by hydride generation and inductively-coupled plasma atomic emission spectrometry. The elements which were found to interfere were Co, Ni, Ag, Au, Bi, Te and Sn. Standard reference materials were an

Determination of arsenic and selenium in
✍ Chris C.Y. Chan; Ram S. Sadana 📂 Article 📅 1992 🏛 Elsevier Science 🌐 English ⚖ 571 KB

Flow-injection analysis is applied to sample introduction in conjunction with automated hydride generation and AAS for the determination of As and Se in environmental samples such as soil, vegetation, waters, sediments, and industrial wastes. A large sample loop was used to provide high sensitivitie

Determination of Ultratrace Selenium and
✍ Lei Wei; Parul Gupta; Robert Hernandez; Fadia Farhat 📂 Article 📅 1999 🏛 Elsevier Science 🌐 English ⚖ 194 KB

Hydride vapor generation atomic absorption spectrometry (AAS) is a preferred method for determination of hydride-forming elements such as selenium and arsenic, particularly at ultratrace concentration levels. In this applied research, the hydride vapor generator-coupled atomic fluorescence spectrosc