MRI compatibility and visibility assessment of implantable medical devices
β Scribed by Beth A. Schueler; Todd B. Parrish; Jyh-Cherng Lin; Bruce E. Hammer; Brian J. Pangrle; E. Russell Ritenour; John Kucharczyk; Charles L. Truwit
- Book ID
- 101312695
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 252 KB
- Volume
- 9
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
β¦ Synopsis
We have developed a protocol to evaluate the magnetic resonance (MR) compatibility of implantable medical devices. The testing protocol consists of the evaluation of magnetic field-induced movement, electric current, heating, image distortion, and device operation. In addition, current induction is evaluated with a finite element analysis simulation technique that models the effect of radiofrequency fields on each device. The protocol has been applied to several implantable infusion pumps and neurostimulators with associated attachments. Experiments were performed using a 1.5-T whole-body MR system with parameters selected to approximate the intended clinical and worst case configuration. The devices exhibited moderate magnetic field-induced deflection and torque but had significant image artifacts. No heating was detected for any of the devices. Pump operation was halted in the magnetic field, but resumed after removed. Exposure to the magnetic field activated some of the neurostimulators.
π SIMILAR VOLUMES
## Abstract Polyimides have been widely used for biosensor encapsulation and more recently as substrates for neural implants. They have excellent thermal stability, high chemical resistance, and can be prepared as thin, flexible films. Photosensitive polyimides present similar physical properties t