MR imaging of patients with implanted drug infusion pumps
β Scribed by Reinhard Von Roemeling; Rosellen M. Lanning; Frederick A. Eames
- Publisher
- John Wiley and Sons
- Year
- 1991
- Tongue
- English
- Weight
- 516 KB
- Volume
- 1
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
β¦ Synopsis
The influence of magnetic resonance (MR) imaging on the integrity and function of an implantable, programmable drug infusion pump and the distortion of the image by this device were tested. Six identical pumps were tested in magnetic fields of 1.5 T. Pump memory was not affected by the procedure. The pump rotor stalled (no infusion) as long as the device was within the magnetic field. Drug delivery resumed as programmed after the pump was removed from the field. No structural damage to the electrical or mechanical pump components was detected. The pump caused a circular image artifact (a signal dropout) within 8-10 cm of the device. The authors conclude that MR imaging is accurate if the area of interest is at least 10 cm from the pump, and that with awareness of temporary infusion cessation, it is safe to perform MR imaging in patients with this implanted pump.
π SIMILAR VOLUMES
## Abstract A 31 year old woman with a 26 year history of Type 1 diabetes mellitus and reslstance to subcutaneous insulin who used a programmable implantable insulin infusion pump for intravenous insulin delivery became pregnant. This report describes our experience of management of this patient th
## Abstract Metallic implants or prostheses can be potentially hazardous during magnetic resonance (MR) imaging because of movement or dislodgment of the foreign object. Magnetic eye implants have been reported to exhibit strong movement when placed in water and exposed to a field of 1.5 T. The aut