MR-diffusion weighted imaging of healthy liver parenchyma: Repeatability and reproducibility of apparent diffusion coefficient measurement
✍ Scribed by Stefano Colagrande; Filippo Pasquinelli; Lorenzo Nicola Mazzoni; Giacomo Belli; Gianni Virgili
- Publisher
- John Wiley and Sons
- Year
- 2010
- Tongue
- English
- Weight
- 317 KB
- Volume
- 31
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose:
To compare repeatability and reproducibility of four different methods of apparent diffusion coefficient (ADC) evaluation of liver parenchyma. In fact, repeatability and reproducibility assessment is mandatory in quantitative evaluations, however, these have not been accurately investigated in liver MR‐diffusion‐weighted studies.
Materials and Methods:
Diffusion‐weighted sequences, b‐value = 0–1000 s/mm^2^, were acquired on 30 healthy volunteers by a 1.5T scanner whose reliability has been validated by a phantom study. Four sampling methods, evaluating various parenchyma percentages by different‐sized region‐of‐interests (ROIs), were compared by two observers: 70% and 30% of the volume, 4%‐one‐ROI‐per‐segment, and 4%‐one‐ROI‐per‐slice in the right‐lobe. Ninety‐five percent limits of agreement and intraclass correlation coefficient (ICC) were calculated.
Results:
Complete measurements on the left lobe could be obtained in less than half of patients. The 4%‐one‐ROI‐per‐slice and 4%‐one‐ROI‐per‐segment yielded lower mean values compared with 30–70% volume methods (1343–1373 versus 1463–1560·10^−6^ mm^2^/s, respectively). Repeatability was acceptable (ICCs ∼ 0.80) whereas reproducibility was low (ICCs ≤ 0.45) for all methods. Averaging at least 3 measurements in middle‐lower sections of the right lobe improved both repeatability (ICCs to ≥0.87) and reproducibility (ICCs to 0.82) for 30–70% V methods.
Conclusion:
ADC measurements were repeatable but not reproducible in our study. Reproducibility could be improved by taking averages on the right lobe with large ROI methods. Studies on procedures that standardize ADC measurements using more than two observers are needed. J. Magn. Reson. Imaging 2010;31:912–920. ©2010 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To determine the utility of diffusion‐weighted MR imaging (DWI) for the diagnosis of adrenal tumors. ## Materials and Methods Forty‐two patients (24 men and 18 women; age, 61.5 ± 12.7 years old; range, 34–86 years) with 43 adrenal tumors (11 functioning cortical adenomas,
## Abstract Magnetic resonance imaging (MRI) remains the standard modality for the local staging of gynecological malignancies but it has several limitations, particularly for lymph node staging or evaluating peritoneal carcinomatosis. Consequently, there has been a growing interest in functional i
## Abstract ## Purpose To evaluate the use of apparent diffusion coefficient (ADC) measurements based on diffusion‐weighted MRI (DWI) to assess stage of liver disease. ## Materials and Methods A total of 31 patients who underwent both a liver biopsy and DWI and 132 patients who only underwent DW
## Abstract ## Purpose To compare single‐shot echo‐planar imaging (SS EPI) diffusion‐weighted MRI (DWI) of abdominal organs between 1.5 Tesla (T) and 3.0T in healthy volunteers in terms of image quality, apparent diffusion coefficient (ADC) values, and ADC reproducibility. ## Materials and Method
## Abstract ## Purpose To evaluate the role of diffusion‐weighted imaging (DWI) in the detection of breast cancers, and to correlate the apparent diffusion coefficient (ADC) value with prognostic factors. ## Materials and Methods Sixty‐seven women with invasive cancer underwent breast MRI. Histo