𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Movement correction of the kidney in dynamic MRI scans using FFT phase difference movement detection

✍ Scribed by E.L.W. Giele; J.A. de Priester; J.A. Blom; J.A. den Boer; J.M.A. van Engelshoven; A. Hasman; M. Geerlings


Book ID
102376244
Publisher
John Wiley and Sons
Year
2001
Tongue
English
Weight
236 KB
Volume
14
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

To measure cortical and medullary MR renograms, regions of interest (ROIs) are placed on the kidney in images acquired using dynamic MRI. Since native kidneys move with breathing, and breath‐holding techniques are not feasible, movement correction is necessary. In this contribution we compare three correction methods, based on image matching, phase difference movement detection (PDMD), and cross‐correlation, respectively. The PDMD‐based method showed the best performance and was able to determine kidney movement in our test series in 68% of the scans with no visible deviation, and in 88% of the scans if a one‐pixel deviation is considered acceptable. J. Magn. Reson. Imaging 2001;14:741–749. © 2001 Wiley‐Liss, Inc.