𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Monolithically integrated μ-capillary electrophoresis with organic light sources and tunable a-Si:H multispectral photodiodes for fluorescence detection

✍ Scribed by Christian Merfort; Konstantin Seibel; Krystian Watty; Markus Böhm


Book ID
104052635
Publisher
Elsevier Science
Year
2010
Tongue
English
Weight
365 KB
Volume
87
Category
Article
ISSN
0167-9317

No coin nor oath required. For personal study only.

✦ Synopsis


The miniaturization of chemical or biochemical analysis systems to micro-or nanoscale dimension allows dramatic acceleration of chemical processes and a high degree of parallelism. In this paper, we present a monolithically integrated application specific lab-on-microchip device (ALM) [1,2] for on-chip capillary electrophoresis (l-CE) with an organic light-emitting diode (OLED) as excitation light source and a tunable a-Si:H multispectral photodiode for colorimetric analysis. The purpose of this work is to adapt the spectral response of a tunable multispectral photodiode to individual fluorescence spectra of different dyes.