Modified heparin-bonded catheter for cannulation of the coronary sinus from the femoral vein

โœ Scribed by Winters, Kenneth J. ;Lasala, John M. ;Eisenberg, Paul R. ;Smith, Stacy C. ;Sewall, David J. ;Shelton, Marc E.


Publisher
John Wiley and Sons
Year
1996
Tongue
English
Weight
519 KB
Volume
39
Category
Article
ISSN
0098-6569

No coin nor oath required. For personal study only.

โœฆ Synopsis


Serial sampling from the coronary sinus is an attractive technique for drawing blood samples to be used in the characterization of procoagulant activity during coronary interventions. We have developed a modlfied Simmons catheter for rapid cannuiation of the coronary sinus from the femoral approach. The catheter design Incorporates heparin bonding and dlstal sideholes to minimize blood sampling artifacts. Coronary sinus cannuiation was performed via the femoral approach in 186 patients by use of a multipurpose catheter (n = a), an unmodified Simmons I or II catheter (n = 64), or the modified Simmons catheter (n = 114). The coronary sinus was cannuiated successfully with the modified Simmons catheter in 97% of patients; the success rate with the unmodified Simmons il catheter was 87% (P = 0.02). The modified Simmons catheter represents an improved technique for cannulation of the coronary sinus from the femoral vein.