Mixing considerations in the development of a gradient microbore high-performance liquid chromatographic system

โœ Scribed by Charles R. Powley; William A. Howard; L.B. Rogers


Book ID
104146748
Publisher
Elsevier Science
Year
1984
Tongue
English
Weight
970 KB
Volume
299
Category
Article
ISSN
1873-3778

No coin nor oath required. For personal study only.

โœฆ Synopsis


A home-made gradient microbore high-performance liquid chromatographic (HPLC) system was designed using two Varian 8500 stepper motor-driven syringe pumps. The pump delivery of the mobile phase solvents in the 10 pl/min flow-rate range is quite pulsed in nature. Static and dynamic mixing devices were evaluated in terms of their ability to homogenize the slugs of solvent delivered by the pumps. Poor mixing was obtained when a simple tee was used to combine the solvents; this incomplete mixing resulted in increased baseline noise, poorer retention time reproducibility, and a significant decrease in c6lumn plate count. The use of static mixing tubes did not yield a completely homogenized mobile phase. A stirred mixing chamber with an internal volume of 80 ~1 gave the best results, and chromatographic performance was not degraded through use of this mixer. Stirred mixing chambers do distort the linearity of a gradient through exponential dilution; a computer simulation of a stirred mixing chamber was developed in order to study the dependency of exponential dilution effects upon mixing-chamber volume, ramp-rate, and flowrate. A simple formula was obtained that allows one to use these variables to calculate the extent of gradient distortion. Finally, comparison of gradient microbore and conventional HPLC for polymer fractionations showed that the two methods performed equally well.


๐Ÿ“œ SIMILAR VOLUMES