Coronary stenting with the half disarticulated Palmaz-Schatz stent is particularly suitable for ostial stenoses, diaphragm stenoses, stenoses distal to tortuous segments or coronary bends and localized dissections after balloon angioplasty. Nevertheless very few data regarding the half stent exist a
Micro stent I, initial results, and six months follow-up by quantitative coronary angiography
โ Scribed by Schalij, Martin J. ;Savalle, Lucas H. ;Tresukosol, Damras ;Jukema, J. Wouter ;Reiber, Johan H. C. ;Bruschke, Albert V. G.
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 73 KB
- Volume
- 43
- Category
- Article
- ISSN
- 0098-6569
No coin nor oath required. For personal study only.
โฆ Synopsis
The Micro stent TM (MS) is a balloon expandable stent that allows the treatment of stenoses in distal and tortuous coronary arteries. This prospective study was performed to evaluate initial and late results of MS implantations. A total of 127 MS (101 in native coronary arteries and 26 in saphenous vein grafts) were implanted in 85 patients (1.5 stents/pt, 65 male, and 20 female, age 62, ุ10 yr) with angina pectoris class II-III: 21 (25%), angina pectoris class IV: 41(48%), and acute myocardial infarction: 23 (27%). Indications per segment treated (nโซ:)39ุโฌ elective: 49 (53%); suboptimal balloon angioplasty (PTCA) result: 33(35%); bailout: 11 (12%). The patients were discharged with 100 mg of aspirin daily unless other indications for oral anticoagulants were present. Procedural success (diameter stenosis of 30% without the occurrence of clinical events within 3 wk) was 85%. Early clinical events (F3 wk included: death:1%; subacute closure: 5%; coronary artery bypass surgery (CABG): 1%; vascular complications: 4%. Late clinical events (3 wk-6 mo) included: acute myocardial infarction:3%, PTCA 5%, CABG 3%, angina class III-IV: 4%. Quantitative angiographic results were: the minimum lumen diameter increased from 0.90 ุ 0.72 before to 3.05 ุ 0.48 mm (FP0.001) after stent implantation. At follow-up, which was 5.5 mo ุ 1.1 mo, 61/79 pts (77%), the loss in diameter was 0.90 ุ 0.68 mm. The net gain was 1.26 ุ 0.90 mm. The restenosis rate (diameter stenosis G 50% at FU) was 13%. This study demonstrates high procedural and late success rates of Micro stent implantations.
๐ SIMILAR VOLUMES
A total of 147 stents were implanted (in overlapping manner in 76% of vessels) in a single coronary artery in 59 patients (60 vessels, 97 lesions, 2.45 stents/vessel) over a period of 18 mo using high pressure stent deployment without ultrasound guidance. The indications for stenting were suboptimal