Tensile and fatigue properties were studied on three kinds of segmented polyether polyurethanes developed for blood pump applications. To study the effects of plasma constituents, specimens were immersed in a cholesterol-lipid solution whose composition is similar to that of plasma. The results obta
Mechanical stability of elastomeric polymers for blood pump applications
β Scribed by Hayashi, Kozaburo ;Takano, Hisateru ;Matsuda, Takehisa ;Umezu, Mitsuo
- Publisher
- John Wiley and Sons
- Year
- 1985
- Tongue
- English
- Weight
- 899 KB
- Volume
- 19
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
β¦ Synopsis
As a series of studies on the mechanical properties of materials used in cardiac prostheses, static and dynamic characteristics and stability of five kinds of elastomeric polymers have been studied by uniaxial tensile and fatigue tests in air at room temperature and in saline solution at 37Β°C. Of all materials tested in this study, Texin MD85A, a segmented polyether polyurethane, has the lowest flexibility under static and dynamic conditions, with relatively high strength. Hexsyn, a polyolefin rubber, is highly flexible with little stress relaxation. However, this material has very low tensile strength and short elongation, and shows unstable change in the elastic modulus during cyclic defor-mation. Avcothane 51, a copolymer of polyurethane and silicon, has unstable mechanical properties and gradually stiffens upon cyclic deformation. On the other hand, Biomer, a segmented polyether polyurethane, has high flexibility and shows the most stable behavior during cyclic deformation regardless of test environment. Toyobo TM5, a similar segmented polyurethane to Biomer, has higher strength and ductility than Biomer, although its static and dynamic flexibility are slightly worse and less stable than those of Biomer. These results indicate that Biomer and Toyobo TM5 are more suitable for flexible components of cardiac prostheses.
π SIMILAR VOLUMES
## Abstract Failure behavior of a polyolefin elastomer which is a candidate material for blood pump applications has been studied under uniaxial and equibiaxial test conditions. Both static and dynamic (fatigue) testing were performed to study four aspects of material failure behavior as suggested
## Abstract **Summary:** Monodisperse polymer microspheres with functional mercapto groups on the surface were prepared from poly[(ethylene glycol dimethacrylate)β__co__β(2βhydroxyethyl methacrylate)], poly(EGDMAβ__co__βHEMA), microspheres successively through esterification of the active hydroxyl