๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Measuring the thickness of articular cartilage from MR images

โœ Scribed by Peter A. Hardy; Premini Nammalwar; Steven Kuo


Publisher
John Wiley and Sons
Year
2001
Tongue
English
Weight
249 KB
Volume
13
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

โœฆ Synopsis


We have developed an algorithm to estimate and display the spatial distribution of the thickness of articular cartilage in human knees. We used a simulation to derive the most appropriate formula for estimating thickness in digital images. The algorithm was tested by imaging a thickness phantom and comparing the results with the known dimensions of the phantom. The results demonstrated that the maximum error encountered in estimating thickness is one pixel while the average error is 0.4 pixels. We imaged a human knee on a cadaver in two separate sessions and used the two image sets to derive thickness maps of the patellar articular cartilage. The thickness maps generated independently from these image sets were very similar, indicating the reliability of the method in deriving accurate thickness estimates. J. Magn. Reson. Imaging 2001;13:120-126.


๐Ÿ“œ SIMILAR VOLUMES


MR imaging of articular cartilage using
โœ Brian A. Hargreaves; Garry E. Gold; Philipp K. Lang; Steven M. Conolly; John M. ๐Ÿ“‚ Article ๐Ÿ“… 1999 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 261 KB ๐Ÿ‘ 1 views