## Abstract A CTβPRESS sequence was implemented on a 3βT MR scanner and optimized for the detection of the C4 resonance of glutamate. By simulating the sequence using the full density matrix it was found that 121 chemical shift encoding steps in __t__~1~ with an increment Ξ__t__~1~ = 1.6 ms were su
Measurement of the point spread function in MRI using constant time imaging
β Scribed by Dr. Matthew D. Robson; John C. Gore; R. Todd Constable
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 840 KB
- Volume
- 38
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
β¦ Synopsis
The point spread function is a fundamental property of magnetic resonance imaging methods that affects image quality and spatial resolution. The point spread function is difficult to measure precisely in magnetic resonance even with the use of carefully designed phantoms, and it is difficult to calculate this function for complex sequences such as echo-planar imaging. This report describes a method that measures the point spread function with high spatial resolution at each pixel in samples of uniform intensity distribution. This method uses additional phase encoding gradients before the echo-planar acquisition that are constant in length but vary in amplitude. The additional gradients are applied to image the contents within each individual voxel. This method has been used to measure the point spread function for echo-planar imaging to demonstrate the effects of limited k-space sampling and transverse relaxation, as well as the effects of object motion. By considering the displacement of the point spread function, local distortions due to susceptibility and chemical shift effects have been quantified and corrected. The method allows rapid assessment of the point spread function in echo-planar imaging, in vivo, and may also be applied to other rapid imaging sequences that can be modified to include these additional phase encoding gradients.
π SIMILAR VOLUMES
The measurement of the velocity of perfused flow in tissue in vivo is complicated by the presence ofstationary as well as moving spins and by a variety ofdephasing phenomena which affect signal amplitudes. This paper describes a technique by which, using flow-sensitized spin-echo sequences with a se
## Abstract Spectroscopic constants of the ground and next seven lowβlying excited states of diatomic molecules CO, N~2~, P~2~, and ScF were computed using the density functional theory SAOP/ATZP model, in conjunction with timeβdependent density functional theory (TDβDFT) and a recently developed S
## Abstract One of the most powerful features of the dynamic contrastβenhanced (DCE) MRI technique is its capability to quantitatively measure the physiological or pathophysiological environments assessed by the passage of contrast agent by means of modelβbased pharmacokinetic analysis. The widely
A novel noninvasive magnetic resonance imaging (MRI) method was developed to determine in vivo blood oxygen saturation and its changes during motor cortex activation in small cerebral veins. Specifically, based on susceptibility measurements in the resting states, pial veins were found to have a mea