Matrix-assisted Laser Desorption/Ionization Using a New Tandem Quadrupole Ion Storage Trap, Time-of-flight Mass Spectrometer
✍ Scribed by Peter Kofel; Markus Stöckli; Jörn Krause; Urs P. Schlunegger
- Publisher
- John Wiley and Sons
- Year
- 1996
- Tongue
- English
- Weight
- 581 KB
- Volume
- 10
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
✦ Synopsis
A tandem quadrupole ion-storage trap, time-of-flight instrument using matrix-assisted laser desorptionl ionization has been constructed. Desorbed ions are lint trapped in an RF' field and collisionally relaxed prior to time-of-!light mass analysis. So far masses up to 90 kDa, a resolution of up to 700 and trapping times of up to 10 minutes have been measured with RF frequencies and amplitudes as low as 20 kHz and 40 V, respectively. The effects of pulse synchronization with the RF voltage and of extraction voltage variation on instrument performance have been investigated. Molecular ions of melittin have been found to undergo reactions in the gas phase during storage.
📜 SIMILAR VOLUMES
## Abstract A prototype matrix‐assisted laser desorption/ionization quadrupole time‐of‐flight (MALDI‐QqTOF) tandem mass spectrometer was used to sequence a series of phosphotyrosine‐, phosphothreonine‐ and phosphoserine‐containing peptides. The high mass resolution and mass accuracy of the instrume
A method for accurate mass measurement of oligonucleotides up to a DNA 35-mer based on matrix-assisted laser desorption/ionization (MALDI) mass spectrometry is described. In this method, a time-lag focusing time-of-flight mass spectrometer is used to achieve high mass resolution to resolve adduct io
The stored waveform inverse Fourier transform (SWIFT) technique is used for broadband excitation of ions in an ion-trap mass spectrometer to perform mass-selective accumulation, isolation, and fragmentation of peptide ions formed by matrix-assisted laser desorptionlionization. Unit mass resolution i
## Abstract We have acquired multi‐stage mass spectra (MS^__n__^) of four branched __N__‐glycans derived from human serum IgG by matrix‐assisted laser desorption/ionization quadrupole ion trap time‐of‐flight mass spectrometry (MALDI‐QIT‐TOF‐MS) in order to demonstrate high sensitivity structural an