๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Material selection in the design of the femoral component of cemented total hip replacements

โœ Scribed by M.J. Fagan; A.J.C. Lee


Publisher
Elsevier Science
Year
1986
Tongue
English
Weight
980 KB
Volume
1
Category
Article
ISSN
0267-6605

No coin nor oath required. For personal study only.

โœฆ Synopsis


The paper is concerned with the selection of materials for use in the design of the femoral component of cemented hip replacement joints. The analysis was performed using a three-dimensional finite element model with a representative physiological loading condition. A range of materials was considered for the stem and cement, enabling the derivation of two equations predicting the maximum critical octahedral shear stress in either material when anv combination of materials is used. The results show that simple substitution of a new material into an existing design could be disastrous, particularly with respect to the cement. For example, a stem manufactured from titanium rather than the usual stainless steel, might lead to an increase in cement stresses of up to 35%. Also, it is shown that carbon fibre reinforced cement should not be used with the basic design of stem considered here.

An enlarged stem was modelled in the same femur, and designed so that, when manufactured from titanium, it had the same flexural stiffness as the basic stem manufactured from stainless steel. The performance of this large stem was also examined in combination with a range of different cement materials.


๐Ÿ“œ SIMILAR VOLUMES


In vitro measurement of strain in the bo
โœ Daniel O. O'Connor; Dennis W. Burke; Murali Jasty; Ronald C. Sedlacek; William H ๐Ÿ“‚ Article ๐Ÿ“… 1996 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 792 KB

## Abstract The strains in the cement mantle surrounding the cemented femoral component of a total hip replacement were measured __in vitro__, using strain gauges embedded within the cement mantle adjacent to the femoral component in femurs from cadavers under physiologic loads simulating both sing

In vitro evaluation of bonding of the ce
โœ Jeff P. Davies; M-K. Tse; W. H. Harris ๐Ÿ“‚ Article ๐Ÿ“… 1995 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 418 KB

Debonding of the cement-metal interface of a cemented femoral component is one mechanism associated with loosening of total hip replacement. An ultrasonic pulse-echo technique was developed to determine, nondestructively, the integrity of the cement-metal interface in vitro. The characteristics of t