In this study a new method is presented for the simultaneous determination of plutonium alpha emitters by liquid scintillation counting using multivariate calibration. The prediction errors obtained for 239+240Pu and '""Pu activities are around 8%. With the procedure proposed, the measurement of the
Liquid scintillation applied to alpha counting of plutonium
β Scribed by Joseph Bubernak
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 457 KB
- Volume
- 276
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
β¦ Synopsis
A liquid scintillation counter has been modified with fast electronic modules to extend its range for a-counting rate measurements. The practical upper limit for accurate counting is about 30 x lo6 cpm, at which 5% dead-time loss corrections are applied. Solutions of ugPu up to 200 g 1-l can be aliquoted directly with capillary pipets, thereby obviating the need for dilutions. The aliquots are mixed with 250 ~1 of cocktail in 400~~1 plastic centrifuge tubes for counting so that waste generation is minimized. Although the counting set-up includes no coincidence circuitry, blanks show less than 10 cpm. Therefore, low-activity samples can also be measured accurately.
π SIMILAR VOLUMES