𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Liquid scintillation applied to alpha counting of plutonium

✍ Scribed by Joseph Bubernak


Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
457 KB
Volume
276
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


A liquid scintillation counter has been modified with fast electronic modules to extend its range for a-counting rate measurements. The practical upper limit for accurate counting is about 30 x lo6 cpm, at which 5% dead-time loss corrections are applied. Solutions of ugPu up to 200 g 1-l can be aliquoted directly with capillary pipets, thereby obviating the need for dilutions. The aliquots are mixed with 250 ~1 of cocktail in 400~~1 plastic centrifuge tubes for counting so that waste generation is minimized. Although the counting set-up includes no coincidence circuitry, blanks show less than 10 cpm. Therefore, low-activity samples can also be measured accurately.


πŸ“œ SIMILAR VOLUMES


Simultaneous determination of plutonium
✍ M. Toribio; J.F. GarcΓ­a; A. Izquierdo-Ridorsa; R. Tauler; G. Rauret πŸ“‚ Article πŸ“… 1995 πŸ› Elsevier Science 🌐 English βš– 740 KB

In this study a new method is presented for the simultaneous determination of plutonium alpha emitters by liquid scintillation counting using multivariate calibration. The prediction errors obtained for 239+240Pu and '""Pu activities are around 8%. With the procedure proposed, the measurement of the