Limitations of clearance determination using the single sample distribution volume method

✍ Scribed by J. Kotzerke; W. Burchert; H. Wiese; U. v. Smekal; H. Hundeshagen


Publisher
Springer
Year
1992
Tongue
English
Weight
481 KB
Volume
19
Category
Article
ISSN
0340-6997

No coin nor oath required. For personal study only.

✦ Synopsis


To investigate the error possibly contained in the single sample distribution volume method for the determination of renal clearance, a mathematical model was applied to describe the effect of changes in distribution volume, clearance, intravascular space, intracompartmental exchange and the time point of blood sampling. The method was found to be valid only under well-defined circumstances (Topt = 45 +/- 5 min, Cl = 390 +/- 50 ml/min, Vd = 16.7 +/- 4 l, alpha = 0.05.V1 ml/min, V1/Vd = 0.5 +/- 0.05) as shown in sample calculations. Two-compartment model-based error calculations demonstrate that this technique implies at best an uncertainty of +/- 10% or more. Whilst it can be used preferably in healthy, normal-weight adults, it is not applicable, without error, under all other circumstances.