Background and

Objective: Laser thrombolysis is a means for clearing blood clots in occluded arteries. Many researchers have studied the mechanisms of clot ablation, and research clinicians have used the technique to treat myocardial infarction with a number of different laser systems. Specifically, a 1-sec pulsed dye laser has been used clinically to remove blood clots in coronary arteries. As a comparative study, the ablation characteristics of lasers with pulse durations in the ranges of 50-150 sec and 2-10 msec were investigated. Two frequencydoubled Nd:YAG lasers at 532 nm were used in this study. Ablation threshold and ablation efficiency of gel phantoms and thrombus using these two lasers were measured and compared with the results of the pulsed dye laser. The pulsed dye laser in this study operated at 522 nm. Study Design/Materials and Methods: Gelatin samples with 150 cm -1 absorption coefficient at 532 nm and animal clot were confined to 3-mm silicone tubes to measure ablation parameters. Additional samples with 150 cm -1 absorption coefficient at 522 nm were prepared for use with the pulsed dye laser. A fluorescence technique and photographic bubble detection were used to determine ablation threshold. A spectrophotometric technique was used to determine ablation efficiency.

Results:

The ablation threshold of the gel phantoms for all three lasers was determined to be 17 ± 2 mJ/mm 2 . Ablation efficiency for the gel phantoms was 1.7 ± 0.1 g/mJ. Clot had an ablation efficiency of 2.9 ± 1.0 g/mJ. Conclusions: Ablation threshold and efficiency are independent of laser pulse duration for 1-sec, 50-150-sec, and 2-10-msec pulses (P < 0.05).


📜 SIMILAR VOLUMES


Laser skin resurfacing using a frequency
✍ Sumian, Chryslain C.; Pitre, Franck B.; Gauthier, B�atrice E.; Bouclier, Martine 📂 Article 📅 1999 🏛 John Wiley and Sons 🌐 English ⚖ 409 KB 👁 1 views

Background and Objectives: Although laser skin resurfacing performed with CO 2 or Er:YAG lasers is efficient, side effects such as prolonged postoperative erythema, delayed healing, scarring, and pigmentation, have been reported. These side effects are due to skin characteristics but also to variati

Treatment of recalcitrant-pigmented flat
✍ Yung-tsai Li; Kao-chia Yang 📂 Article 📅 2001 🏛 John Wiley and Sons 🌐 English ⚖ 163 KB

## Abstract ## Background and Objective Recalcitrant‐pigmented flat warts may persist for years despite treatment. Many methods have been used to manage pigmented flat warts, but their efficiencies are varied. Since pigmented flat warts are very similar to lentigo clinically, we used frequency‐dou

Are copper vapour and frequency doubled
✍ Sheehan-Dare, Robert A.; Cotterill, John A. 📂 Article 📅 1996 🏛 John Wiley and Sons 🌐 English ⚖ 508 KB

## Background and objective: The copper vapour laser (CVL) and the frequency doubled Nd:YAG laser (FDNYL) have been increasingly adopted as alternatives to the argon laser for treating vascular skin lesions. Automated scanning devices that permit the use of any of these lasers at 3&50 ms pulse wid

Double-frequency Nd:YAG laser vs. argon-
✍ Bandello, Francesco; Brancato, Rosario; Lattanzio, Rosangela; Trabucchi, Giusepp 📂 Article 📅 1996 🏛 John Wiley and Sons 🌐 English ⚖ 403 KB 👁 2 views

## Background and objective: A randomized clinical trial using the argon-green (514 nm) and the double-frequency nd:yag (532 nm) lasers was carried out on 42 eyes with proliferative diabetic retinopathy (pdr), with the aim of assessing the long-term effects of double-frequency nd:yag panretinal pho

Single-treatment skin tightening by radi
✍ Douglas J. Key 📂 Article 📅 2007 🏛 John Wiley and Sons 🌐 English ⚖ 279 KB

## Abstract ## Background and Objective To compare single‐treatment facial skin tightening achieved with the current radiofrequency (RF) protocol with single‐treatment tightening achieved with the long‐pulsed, 1064‐nm Nd:YAG laser. ## Study Design/Materials and Methods A total of 12 patients wer