𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Laser doppler monitoring of replants using a small prism probe

✍ Scribed by Ian M. R. Lowdon; E. Bruce Toby; Jeff Ecker; Anthony V. Seaber; James R. Urbaniak


Book ID
102513543
Publisher
John Wiley and Sons
Year
1989
Tongue
English
Weight
251 KB
Volume
10
Category
Article
ISSN
0738-1085

No coin nor oath required. For personal study only.

✦ Synopsis


Eighteen patients admitted to Duke University Medical Center for upper limb replantation or revascularization were studied. A small laser Doppler flow prism probe and a cutaneous temperature probe were attached to the distal pulp of replanted digits. Measurements were recorded hourly for the first 3 postoperative days. In the successful cases, the mean laser Doppler flow measurement was 1.76 units (lower 97% confidence limit 0.5 units) and the mean temperature was 34.1"C (lower 97% confidence limit 32Β°C). In 4 patients with vascular compromise, the laser Doppler flow correctly identified the problem. The reduced size of the new laser Doppler probe makes monitoring of distal pulp flow technically easier while maintaining accuracy.


πŸ“œ SIMILAR VOLUMES


Small-probe-volume laser Doppler anemome
✍ S.R. Martin; C.J. Bates πŸ“‚ Article πŸ“… 1992 πŸ› Elsevier Science 🌐 English βš– 639 KB

A special-purpose high-resolution laser Doppler optical system has been used to investigate the wall region of a turbulent water flow within a rectangular channel. The exceptionally small control volume (30 I~m x 17 I~m x 15 pom) has enabled unique LDA measurements to be achieved on two walls of the