The ion optics of multi-turn time-of-Γight (TOF) mass spectrometry was investigated. Two feasible systems using toroidal sector plus quadrupole lenses and cylindrical sectors plus quadrupole doublets were found. The combination of a linear TOF with a Γgure-of-eight type multi-turn system is proposed
An external source 7 T Fourier transform ion cyclotron resonance (FTICR) mass spectrometer offers three main novel features. First, a 9-way ion-source cross allows for mounting of up to three ionization sources simultaneously, thereby minimizing 'downtime' for changing ion sources. Second, an electr