๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Investigation of getter-ion pump operation in systems containing oil and mercury vapour

โœ Scribed by BJ Ward; JC Bill


Publisher
Elsevier Science
Year
1966
Tongue
English
Weight
354 KB
Volume
16
Category
Article
ISSN
0042-207X

No coin nor oath required. For personal study only.

โœฆ Synopsis


Details are given of some attempts to start a getter-ion pump which was heavily contaminated with oil vapour. Baking the pump while it was maintained under vacuum and the use of a liquid nitrogen cold trap are two methods of removing oil contamination and restoring pump operation. The effects of mercury vapour on an 8 litre/sec getter-ion pump are investigated. The getter-ion pump failed to operate with mercury vapour present at room temperature, however, in mercury vapour atmospheres less than 1 ร— 10 -s torr the getter-ion pump worked satisfactorily. Continuous operation of the pump was possible with vapour pressures below 1 ร— 10 -7 torr without any apparent deterioration in pump performance.