𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Investigation of flow and material induced hemolysis with a Couette type high shear system

✍ Scribed by S. Klaus; R. Paul; H. Reul; K. Mottaghy; B. Glasmacher


Publisher
John Wiley and Sons
Year
2001
Tongue
English
Weight
160 KB
Volume
32
Category
Article
ISSN
0933-5137

No coin nor oath required. For personal study only.

✦ Synopsis


Damage of red blood cells (hemolysis) in miniaturized pump systems for heart support is induced by contact with artificial surfaces and high mechanical shear forces. In vitro experiments with porcine blood under well defined material and flow conditions with a new Couette model showed hemolysis not starting until shear stresses of 400 Pa and exposure times of 400 ms. Hemolysis in general was much lower than predicted in earlier investigations. Heparinized blood revealed a more sensitive behaviour as compared to citrated blood (CPDA-1).

Die Scha Èdigung roter Blutko Èrperchen (Ha Èmolyse) in miniaturisierten Pumpsystemen zur Herzunterstu Ètzung wird durch Kontakt zu