Investigating ion-surface collisions with a niobium superconducting tunnel junction detector in a time-of-flight mass spectrometer
✍ Scribed by G. Westmacott; F. Zhong; M. Frank; S. Friedrich; S. E. Labov; W. H. Benner
- Publisher
- John Wiley and Sons
- Year
- 2000
- Tongue
- English
- Weight
- 125 KB
- Volume
- 14
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
✦ Synopsis
The performance of an energy sensitive, niobium superconducting tunnel junction (STJ) detector is investigated by measuring the pulse height produced by impacting molecular and atomic ions at different kinetic energies. Ions are produced by laser desorption and matrix-assisted laser desorption in a time-offlight mass spectrometer. Our results show that the STJ detector pulse height decreases for increasing molecular ion mass, passes through a minimum at around 2000 Da, and then increases with increasing mass of molecular ions above 2000 Da. The detector does not show a decline in sensitivity for high mass ions as is observed with microchannel plate ion detectors. These detector plus height measurements are discussed in terms of several physical mechanisms involved in an ion-surface collision.