𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Intracranial pressure monitoring in children: comparison of external ventricular device with the fiberoptic system

✍ Scribed by G. Gambardella; C. Zaccone; E. Cardia; F. Tomasello


Publisher
Springer
Year
1993
Tongue
English
Weight
392 KB
Volume
9
Category
Article
ISSN
0256-7040

No coin nor oath required. For personal study only.

✦ Synopsis


Several intracranial pressure monitoring devices have been developed in the past several years. We have recently adopted the Camino fiberoptic device that permits subdural, intraparenchymal, and intraventricufar monitoring. In this report we compare experiences in monitoring a group of pediatric patients with severe craniocerebral trauma and coma, grouped according to severity of Glasgow Coma Scale score. Patient age ranged from 2 to 16 years. Twelve patients were monitored by a ventricular catheter and 37, treated more recently, by a Camino fiberoptic device. The study demonstrated that the fiberoptic device and the ventricular catheter have the same accuracy and reliability. The fiberoptic method correlates very closely with the ventriculostomy method, but the pressure values are always 3 + 2 mmHg lower than those obtained with the conventional pressure transducer system, especially in more critically ill patients. This new technique is also easier to implant, safer to use, has minimal drift, and is minimally invasive, which particularly speaks for its use in pediatric patients.