Initial clinical experience with a new MRI vacuum-assisted breast biopsy device

✍ Scribed by Mary C. Mahoney


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
127 KB
Volume
28
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To report initial clinical experience with a new previously unreported vacuum‐assisted device (EnCor MR, SenoRx, Aliso Viejo, CA) for magnetic resonance imaging (MRI)‐guided breast biopsy and to compare this device with previously reported technologies.

Materials and Methods

A prospective study of MRI‐guided vacuum‐assisted breast biopsies using the EnCor device was undertaken. Lesions targeted for biopsy were occult on mammography, ultrasound, and physical examination. Histologic results were obtained for all cases. Patients with benign, concordant results underwent clinical and imaging follow‐up. Cases with discordant, high‐risk, or malignant results were referred for surgical biopsy.

Results

A total of 55 biopsies were performed in 47 patients. All biopsies were technically successful. Histologic results yielded malignancy in 10 (10/55, 18%) of the 55 lesions and high‐risk or atypical lesions in 7 (7/55, 13%) of the cases. All of these patients have had surgical follow‐up, with four (4/55, 7%) underestimations. There were three (3/55, 5%) discordant cases. 35 (35/55, 64%) lesions yielded benign concordant results. Of these, six (6/35, 17%) lesions were removed surgically, with confirmation of the MR‐guided biopsy results. Twenty (20/35, 57%) lesions have undergone at least 6‐month follow‐up MRI, with resolution or stability of the MRI findings.

Conclusion

The EnCor breast biopsy device has comparable results to those previously reported for other MRI‐guided vacuum‐assisted technologies. J. Magn. Reson. Imaging 2008;28:900–905. © 2008 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


An MRI-compatible semiautomated vacuum-a
✍ Bruce L. Daniel; Lynetta J. Freeman; Jessica M. Pyzoha; Terry D. McCoy; Robyn L. 📂 Article 📅 2005 🏛 John Wiley and Sons 🌐 English ⚖ 761 KB

## Abstract ## Purpose To develop an MR‐compatible vacuum‐assisted core needle breast biopsy system. ## Materials and Methods A vacuum‐assisted breast biopsy system (Mammotome Hand Held; Ethicon Endo‐Surgery, USA) was modified for freehand MRI‐guided biopsy in an open, interventional 0.5‐T scann

Rotational atherectomy with a new device
✍ Kjellgren, Olle ;Motarjeme, Amir ;Feld, Steven ;Mishkel, David C. ;Underwood, Ca 📂 Article 📅 1996 🏛 John Wiley and Sons 🌐 English ⚖ 814 KB

The Bard Atherectomy Catheter is a new rotational atherectomy device that consists of a flexible, hollow, thin-walled cutting catheter that, while rotated at 1,500 revolutions per minute, is advanced across the lesion over a special spiral guidewire system. We report the initial clinical experience