β¦ Synopsis
There are limitations to all techniques used to identify particulate polyethylene in histological specimens. The goal of our study was to determine if remnant metal elements used during the catalytic production of ultra high molecular weight polyethylene, could be used as markers for particulate polyethylene detection in histological specimens. It was hypothesized that these catalyst elements could be detected in polyethylene using energy dispersive X-ray elemental analysis. Six samples from five different companies were evaluated. These included virgin polymer powder, polyethylene bar stock, and artificial joint components.
Five specimens from each of the six samples were analyzed with energy dispersive X-ray elemental analysis. After elemental analysis was completed, only 2 of 30 specimens were positive for the catalyst elements. In the remaining 28 specimens, catalyst elements were not detected. Our investigation demonstrates that energy dispersive X-ray elemental analysis is not currently a feasible method of particulate polyethylene detection. Additional techniques will need to be developed to accurately identify particulate polyethylene in histological specimens.
π SIMILAR VOLUMES
Energy-dispersive x-ray Γuorescence (EDXRF) was applied to the analysis of environmental samples, such as algae, microalgae and marine sediments, prepared as thin samples by mixing powder with two-component epoxy resins, and as intermediate samples in the form of pellets. Further, sulphur, the conce
Performance characteristics of a microstrip gas counter operated as an x-ray Γuorescence spectrometer are reported. Gas ampliΓcation as a function of microstrip anode-cathode voltage was measured, and the breakdown threshold voltage was determined in pure xenon. The detector temporal stability and t
Energy-dispersive x-ray Γuorescence in combination with partial least-squares regression (PLSR) was used to determine nickel, copper and arsenic in aqueous solutions. Di β erent spectrum pre-processing methods (square root and logarithmic transformation and top hat Γltering) were used prior to the PL
We have developed a rapid, nondestructive technique using energy-dispersive X-ray fluorescence spectroscopy to determine trace element concentrations in samples of archaeological felsite. The method employs a series of secondary targets that are selected to provide optimal sensitivity to groups of e