In Vivo Validation of Magnetic Resonance Blood Volume Flow Measurements with Limited Spatial Resolution in Small Vessels

✍ Scribed by Mark B. M. Hofman; Frans C. Visser; Albert C. Van Rossum; Ger Q. M. Vink; Michiel Sprenger; Nico Westerhof


Publisher
John Wiley and Sons
Year
1995
Tongue
English
Weight
844 KB
Volume
33
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The accuracy of magnetic resonance phase contrast volume flow measurements in small blood vessels is expected to be smaller than in large vessels, because of partial volume effects at the vessel boundary. Accuracy was validated in the dog femoral artery, diameter 3.5 ± 0.7 mm, using an ultrasonic transit‐time flowmeter (TT). The number of pixels per vessel diameter (N~D~) ranged from 1.6 to 4.8. The vessel cross‐section was determined using a threshold in the magnitude image. Between the two methods the correlation coefficient was 0.95 (range 10–200 ml/min). The proportional difference (PD), (Q~TT~ – Q~MR~/1/2(Q~TT~ + Q~MR~), was 0.8%, showing no systematic difference between the methods. The PDs standard deviation was 27%, and 19% for flow rates above 30 ml/min. Only a significant decrease of the PDs variance was found at the highest N~D~ values, suggesting other sources of error than partial volume effects. It is concluded that with an N~D~ value of about 3, accurate blood volume flow rates can be determined.