In vivo evaluation of modified mandrel-grown vascular prostheses

โœ Scribed by Werkmeister, Jerome A. ;Edwards, Glenn A. ;White, Jacinta F. ;Casagranda, Franca ;Hunt, John A. ;Williams, David F. ;Ramshaw, John A. M.


Book ID
102654521
Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
577 KB
Volume
47
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

โœฆ Synopsis


The Omniflowโ„ข Vascular Prosthesis (OVP) has been manufactured and extensively tested in animal and human trials. It has mechanical and biological qualities superior to synthetic and biological conduits, particularly in low flow conditions. For further development into the smaller diameter coronary prostheses, the inner luminal surface is of paramount importance. In a previous study this inner surface was modified to produce a more uniformly thicker nonundulating surface. In this study the mandrels of these modified OVPs were treated with either collagen or heparin; the OVPs were evaluated for patency, tissue integration and wound healing, and endothelialization using a dog model comparable to that used to evaluate the unmodified OVP. In all instances, each of the modified prostheses were fully patent and had no signs of any deleterious effects caused by these modifications; no thrombus or aneurysms were visible. The tissue response was rapid with excellent new host collagen deposition within the vessel wall and minimal inflammatory and foreign body giant cells. Endothelialization was noted at the earliest explant time point in central regions of the prostheses, albeit that the histological picture at this time point appeared to reflect a complex atypical intimal layer.


๐Ÿ“œ SIMILAR VOLUMES


In vivo evaluation of a biomimetic apati
โœ Vasudev, Deepta Vani ;Ricci, John L. ;Sabatino, Christopher ;Li, Panjian ;Parson ๐Ÿ“‚ Article ๐Ÿ“… 2004 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 277 KB

## Abstract Osteoconductive mineral coatings represent an established technology for enhancing the integration of orthopedic implants with living bone. However, current coatings have limitations related to fabrication methods, attachment strength to metal substrates, and in vivo performance. Low te