An in vivo investigation of a new high molybdenum and nitrogen duplex stainless steel (25Cr-7Ni-4Mo-00.3N) has been performed. Cylindrical pins and specially developed devices, to test in static conditions the in vivo localized corrosion resistance, made of this new duplex steel and of a common aust
In vivo evaluation of a high-strength, high-ductility stainless steel for use in surgical implants
✍ Scribed by Syrett, Barry C. ;Davis, Edward E.
- Publisher
- John Wiley and Sons
- Year
- 1979
- Tongue
- English
- Weight
- 1006 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A high‐strength, high‐ductility, austenitic stainless steel has been evaluated for use in surgical implants by performing in vivo tests in rats, rabbits, dogs, and rhesus monkeys. This stainless steel, a TRIP (TRansformation Induced Plasticity) steel containing about 4% Mo, was compared with two alloys in current clinical use: Type 316L stainless steel and cast Vitallium. Compared with the other two alloys, cast Vitallium generally had higher resistance to corrosion and superior biocompatibility in all animals. The tests in rats and dogs indicated that the corrosion resistances of the TRIP steel and the Type 316L stainless steel were similar and that the tissue reactions caused by these alloys were also similar. However, in rhesus monkeys, the TRIP steel was shown to be susceptible to stress‐corrosion cracking and much more susceptible to crevice corrosion than Type 316L stainless steel. Limited tests in rabbits supported the observation that the TRIP steel is susceptible to stress‐corrosion cracking. These inconsistencies in the in vivo tests underline the need for a reevaluation of the popular test techniques and of the animals commonly chosen for assessing the suitability of candidate implant materials. The “worst case” results from the rhesus monkey tests were entirely consistent with previous results obtained from in vitro studies. However, further work must be performed before the behavior of metals in humans, rhesus monkeys, or any other animal, can be predicted with confidence from an in vitro test program.
📜 SIMILAR VOLUMES
## Abstract Experience with prestressed concrete over about half a century has indicated that the corrosion resistance of conventional prestressing steel does not always satisfy, especially the prestressing steels are susceptible to chloride attack (de‐icing salts) and hydrogen (hydrogen‐induced st
## Abstract Detailed examination has been carried out on two specimens of 18/8 steels representing the extremes of the rates of parabolic duplex scale growth at 600°C using metallography, scanning electron microscopy, microprobe and nuclear microprobe analysis. The high growth rate material, vacuu
## Abstract ## Purpose To assess magnetic resonance (MR) pulse sequences for high resolution intravascular imaging. ## Materials and Methods Intravascular imaging of the abdominal aorta and iliac arteries was performed in vivo in a porcine model at 1.5 T using catheter‐mounted micro‐receive coil
was evaluated for in vivo brain proton surface coil NMR spectroscopy. The studies were performed on cat brain using surface coils at 4.7 T. The sequence produced brain spectra with adequate water suppression, and a broader excitation profile than sequences which form spin echoes using 133 1 pulses (