A point-resolved spectroscopy (PRESS)-localized double quantum filter was implemented on a 1.5T clinical scanner for the estimation of gamma-amino butyric acid (GABA) concentrations in vivo. Several calibrations were found to be necessary for consistent results to be obtained. The apparent filter yi
In vivo detection of GABA in human brain using a localized double-quantum filter technique
✍ Scribed by John R. Keltner; Lawrence L. Wald; Blaise De B. Frederick; Perry F. Renshaw
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 495 KB
- Volume
- 37
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A proton MR spectral editing technique employing a spatially localized, double‐quantum filter (DQF) was used to measure γ‐aminobutyric acid (GABA) in the human brain at 1.5 T. The double‐quantum method provided robust, single‐shot suppression of uncoupled resonances from choline, creatine, and NAA and allowed detection of the γCH~2~ GABA (3.0 ppm) resonance with 30% efficiency. Spatial localization of the GABA measurement was achieved by incorporating PRESS localization within the double‐quantum excitation and detection sequence. A calibration technique was developed to adjust the relative phases of the RF pulses to maximize the in vivo double‐quantum detection efficiency for an arbitrary voxel location. The sequence efficiency, degree of suppression of uncoupled resonances, and characterization of the in vivo DQF technique was examined in phantom experiments and in a study of the occipital lobe of 10 normal subjects. The ratio of the 3.0‐ppm GABA resonance to the 3.0‐ppm creatine resonance was found to be 0.20 ± 0.05 (SD).
📜 SIMILAR VOLUMES
## Abstract In this work, a new double quantum filter for glutathione (GSH) editing is introduced, combined with the point‐resolved spectroscopy localization sequence (PRESS), and demonstrated in vivo. Compared to the conventional double quantum coherence filter, the new filter has two major advant
## Abstract An improved double quantum coherence (DQC) filter for the selective in vivo detection of glutathione (GSH) in the human brain at 1.5 Tesla is presented. The goal was to minimize contamination of the DQC‐filtered GSH signal at 2.9 ppm with contributions arising from GABA. The modificatio
## Abstract A method is described to selectively measure lactate __in vivo__ using proton zero quantum coherence techniques. The signal from lipids is eliminated. A surface coil and additionally slice selective localization are used. The resulting spectra demonstrate the good performance of the met