In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: Precision, accuracy, and signal-to-noise ratios

✍ Scribed by Joachim Lotz; Rolf Döker; Ralph Noeske; Meike Schüttert; Roland Felix; Michael Galanski; Matthias Gutberlet; Gerd Peter Meyer


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
327 KB
Volume
21
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To evaluate the signal‐to‐noise ratio (SNR), precision, and accuracy of phase‐contrast flow measurements at 3 T with the help of an in vitro model and to compare the results with data from two 1.5‐T scanners.

Materials and Methods

Using an identical setup of a laminar flow model and sequence parameters, measurements were done at one 3‐T and at two 1.5‐T systems. Precision, accuracy, and SNR were obtained for velocity encodings ranging from 55 up to 550 cm^−1^. SNRs were calculated from the magnitude as well as the flow encoded images.

Results

Precision and accuracy for the in vitro flow model were similarly high in all scanners with no significant difference. For velocity encodings from 55 cm^−1^ up to 550 cm^−1^, the SNR in magnitude as well as phase encoded images of the 3‐T measurements was approximately 2.5 times higher than the SNR obtained from the two 1.5‐T systems.

Conclusion

Even without optimization for the 3‐T environment, flow measurements show the same high accuracy and precision as is known from clinical 1.5‐T scanners. The superior SNR at 3 T will allow further improvements in temporal and spatial resolution. This will be of interest for small‐size vessels like coronary arteries or for slow diastolic flow patterns. J. Magn. Reson. Imaging 2005;21:604–610. © 2005 Wiley‐Liss, Inc.