With packed and capillary on-column experiment, the C.V. value improved. The C.V. differences for these trials may be due to the different injection temperatures. Although these C.V. values are larger than those reported elsewhere [3], this may relate to the
Improved capillary GC/MS interface
β Scribed by Giang, B. Y.
- Book ID
- 102349055
- Publisher
- John Wiley and Sons
- Year
- 1984
- Tongue
- English
- Weight
- 210 KB
- Volume
- 7
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
π SIMILAR VOLUMES
## Abstract This paper describes a design for an allβglass GC/MS interface. The design has advantages due to low dead volume, lack of column effluent adsorption and degradation, simple interface production and installation, and ease of column mounting/dismounting.
## Abstract A new allβglass interface for GC/MS designed to fully exploit the potential of glass capillary columns is described. The system works in the open split mode and has high versatility with respect to column changing. It is especially suitable for mass spectrometers, which are not exclusiv
## Mass spectrum of the component C54 (see Flg. 1) Flgure 2 Mass spectrum of the component C54 (see Flg. 1)