Impact of heart rate variability in patients with normal sinus rhythm on image quality in coronary magnetic angiography

✍ Scribed by Tarinee Tangcharoen; Cosima Jahnke; Uwe Koehler; Bernhard Schnackenburg; Christoph Klein; Eckart Fleck; Eike Nagel


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
248 KB
Volume
28
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To evaluate the influence of heart rate variability on image quality in patients with suspected coronary artery disease during magnetic resonance coronary angiography (MRCA).

Materials and Methods

The coronary images from 70 patients who underwent target‐volume MRA were retrospectively analyzed. Two independent observers evaluated image quality using a score from 0 (nonvisible) to 4 (excellent quality). Images were grouped into good (score > 2) and poor image quality (score 0–2). Five parameters (effective scan duration, navigator efficiency, mean heart rate, acquisition window, and heart rate variability) were evaluated.

Results

In all, 56 of the 70 patients (80%) were scored as good and 14 as poor image quality. Only the navigator efficiency, heart rate variability, and acquisition window showed statistically significant for the prediction of image quality, with navigator efficiency being the strongest predictor. Subgroup analysis showed that patients with navigator efficiency >60% had good image quality independent of the heart rate variability. In patients with navigator efficiency ≤60%, heart rate variability is the only remaining significant parameter (P = 0.03).

Conclusion

Navigator efficiency was the most important predictor of MRCA image quality. Heart rate variability is the most important predictor of image quality in patients with a navigator efficiency ≤60%. J. Magn. Reson. Imaging 2008;28:74–79. © 2008 Wiley‐Liss, Inc.