## Abstract ## Purpose To develop a dedicated radiofrequency (RF) coil for high‐resolution magnetic resonance imaging (MRI) of finger joints at 3T to improve diagnostic evaluation of arthritic diseases. ## Materials and Methods A dedicated cylindrical RF receive coil was developed for imaging fi
Imaging of finger joints in a whole-body MR system using a simple and low-cost solenoidal coil
✍ Scribed by Helmut Rumpel; Wilfred C.G. Peh
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 162 KB
- Volume
- 14
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
This work considers the design and operation of a solenoid radiofrequency (RF) coil for magnetic resonance (MR) mini‐imaging of finger joints using a whole‐body system. The considerations were focused on the requirements of high sensitivity, easy patient access, and cost‐effectiveness. The proposed design is a short, two‐turn solenoidal coil that accomplishes high spatial uniformity over a 2‐cm region of interest (ROI). The coil resembles an oversized ring, and easily fits fingers that are spread. An experimental demonstration of the proposed approach is given by imaging the proximal interphalangeal joint in healthy volunteers using a 2.5‐cm field of view (FOV). Spin‐echo and gradient‐recalled‐echo T1‐weighted sequences, and a 3D spoiled‐gradient‐echo sequence were used. J. Magn. Reson. Imaging 2001;14:800–802. © 2001 Wiley‐Liss, Inc.