High-voltage capillary zone electrophoresis of red blood cells

โœ Scribed by An Zhu; Yi Chen


Publisher
Elsevier Science
Year
1989
Tongue
English
Weight
580 KB
Volume
470
Category
Article
ISSN
1873-3778

No coin nor oath required. For personal study only.

โœฆ Synopsis


The high-voltage wide-bore capillary zone electrophoresis of red blood cells was investigated. The reproducibility of the retention time (electrophoretic mobility) is excellent and the differentiation among various species is good. The peaks in the electropherogram describe the distribution of the size and/or surface charge of the cells and are therefore broad. The relationship between the peak height and the number of cells injected is good, with linear correlation coefficients better than 0.98. Details of the preparation of cell suspensions and support electrolytes are given, which is essential for obtaining reproducible results. The inner surface of FEP capillary tubing is degraded by the application of high voltage and a pause is necessary between successive experiments if good and reproducible peak shapes are to be obtained. The length of the pause increases with the number of experiments made, and finally the tubing becomes useless. Inspection of the inner surface by X-ray photoelectron spectroscopy showed the breakdown of CHF bonds, but the actual mechanism is not known.