High-resolution renal MRA: Comparison of image quality and vessel depiction with different parallel imaging acceleration factors
✍ Scribed by Henrik J. Michaely; Karin A. Herrmann; Harald Kramer; Olaf Dietrich; Gerhard Laub; Maximilian F. Reiser; Stefan O. Schoenberg
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 361 KB
- Volume
- 24
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To investigate the image quality and vessel depiction of renal MRA with integrated parallel imaging techniques (iPAT) using acceleration factors of 2 and 3.
Materials and Methods
In this prospective study renal MRA was performed on 14 and 12 patients with acceleration factors of 3 and 2, respectively. For the MRA a 3D‐GRE sequence with an acquired spatial resolution of 0.9 × 0.8 × 1.0 mm^3^ was applied (TR/TE = 3.79 msec/1.39 msec, FOV = 400 mm × 320 mm, acquired matrix = 512 × 384, flip angle = 25°) on a 32‐channel 1.5T MR scanner. The acquisition time was 26 seconds with iPAT 2, and 19 seconds with iPAT3. All parameters other than acquisition time and acceleration factor were kept constant. To assess the signal‐to‐noise ratio (SNR) we performed repetitive phantom measurements using iPAT 2 and 3. The images were rated by two radiologists in terms of noise, artifacts, and the quality of vessel depiction for the proximal, segmental, and subsegmental renal artery. A Mann‐Whitney U‐test and kappa‐test were used for statistical analysis.
Results
SNR decreased significantly with iPAT 3 in the phantom measurements. The two readers found no difference in noise, but significantly fewer artifacts with iPAT 3. The depiction of segmental vessels was significantly better for both readers with iPAT 3, and the subsegmental vessels were rated significantly better by one reader. iPAT 3 also resulted in a better interreader agreement.
Conclusion
The use of iPAT 3 for renal MRA enables a better depiction of the distal parts of the renal artery. The decrease in SNR is not diagnostically impairing. J. Magn. Reson. Imaging 2006. © 2006 Wiley‐Liss, Inc.