High precision trace analysis in capillary gas-liquid chromatography using an inlet based on the static solvent effect

✍ Scribed by Lawson, K. H. ;Pretorius, V. ;Apps, P. J.


Publisher
John Wiley and Sons
Year
1987
Tongue
English
Weight
365 KB
Volume
10
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

✦ Synopsis


A sample inlet, based on the static solvent effect has been designed and constructed. The inlet can handle samples up to 20 1-11 which allows for the detection, using an F.I.D., of constituents in the parts per billion range. The reproducibility of the inlet in terms of peak percentage area is around 1% relative standard deviation over a range of chemical types.