Hematopoietic progenitor cell large volume leukapheresis (LVL) on the Fenwal Amicus blood separator
β Scribed by Edwin A. Burgstaler; Alvaro A. Pineda; Jeffrey L. Winters
- Book ID
- 102299469
- Publisher
- John Wiley and Sons
- Year
- 2004
- Tongue
- English
- Weight
- 144 KB
- Volume
- 19
- Category
- Article
- ISSN
- 0733-2459
No coin nor oath required. For personal study only.
β¦ Synopsis
A technique for large volume leukapheresis (LVL) for hematopoietic progenitor cell (HPC) collection using the Fenwal Amicus is presented. It was compared to standard collections (STD) with regard to CD34+ cell yields and cross-cellular content. Optimal cycle volumes and machine settings were evaluated for LVL procedures. A total of 68 patients underwent 80 HPC collection procedures. Because of differences in CD34+ cell yields associated with peripheral white blood cell counts (WBC), the comparison was divided into groups of 20 with WBC < or =35 x 10(9)/L (< or =35 K) and those >35 x 10(9)/L (>35 K). Baseline CD34+ cell counts (peripheral count when patient started HPC collection) were used (median 18-23 cells/microl). Significantly more whole blood (corrected for anticoagulant) was processed with LVL (LVL 20 l vs. STD 13.5 l). For < or =35 K, median CD34+ x 10(6), WBC x 10(9), RBC ml, Plt x 10(11) yields/collection were 183, 21.2, 14, 0.8, respectively, for STD vs. 307, 22.1, 11, 1.0, respectively, for LVL. For >35 K, median CD34+ x 10(6), WBC x 10(9), RBC ml, Plt x 10(11) yields/collection were 189, 32.7, 15, 1.4, respectively, for STD vs. 69, 40.8, 21, 1.3, respectively, for LVL. We have described a method of LVL using the Amicus that, in patients with pre-procedure WBC < or =35 x 10(9)/L, collects more CD34+ cells than a standard procedure with acceptable cross-cellular content. This method is not recommended when pre-procedure WBC counts are >35 x 10(9)/L.
π SIMILAR VOLUMES
## Abstract A technique was developed to improve consistency of MNC transfers from the centrifuge to the collection bag in the Fenwal Amicus. The operator assures that RBCs completely fill the cassette by the end of the transfer by adjusting the RBC offset in succeeding cycles. We compared yields a
## Abstract The current paper reports on the predicting factors associated with satisfactory peripheral blood stem cell collection and the efficacy of largeβvolume leukapheresis (LVL) using femoral vein catheterization to harvest PBSCs with Fenwal CS 3000 Plus from normal healthy donors for allogen