We have used a simple affinity chromatography method to measure total glycosylated haemoglobins and glycosylated plasma proteins in maternal and cord blood at 50 normal deliveries. The affinity method gives equal weighting to glycosylated haemoglobins including haemoglobin F in cord blood. The mean
Glycosylated haemoglobin and plasma glycoprotein assays by affinity chromatography
โ Scribed by D. G. Willey; M. A. Rosenthal; S. Caldwell
- Publisher
- Springer
- Year
- 1984
- Tongue
- English
- Weight
- 275 KB
- Volume
- 27
- Category
- Article
- ISSN
- 0012-186X
No coin nor oath required. For personal study only.
โฆ Synopsis
A rapid chromatography method for separating glycosylated protein from non-glycosylated protein, using a boronate-agarose affinity medium which selectively binds the cis-diol groups of glycoproteins, was used to quantitate plasma glycoprotein as well as glycohaemoglobin. The results were found to be independent of: (1) temperature from 16.5 to 29.8 degrees C; (2) haemoglobin variants, and (3) aldimine glycoprotein adducts. Thus several of the common problems occurring in existing haemoglobin A1 assays are eliminated. There was a close correlation between glycohaemoglobin measured by affinity chromatography and haemoglobin A1 by cation-exchange (r = 0.959, n = 215). Specimens from 103 diabetic patients and 112 healthy volunteers were assayed. The following reference ranges were established: glycohaemoglobin 5.5-8.4%; glycosylated total protein 11.5-16.2%; glycoalbumin 11.6-19.5%.
๐ SIMILAR VOLUMES
## Abstract Reproducible and efficient affinity enrichment is increasingly viewed as an essential step in many investigations leading to the discovery of new biomarkers. In this work, we have evaluated the repeatability of lectin enrichment of glycoproteins from human blood serum through both quali
## Abstract Interactions of the drug carbamazepine with the serum protein ฮฑ~1~โacid glycoprotein (AGP) were examined by highโperformance affinity chromatography. Frontal analysis studies with an immobilized AGP column and control column indicated carbamazepine had both lowโaffinity interactions wit
## Abstract A new group of specificโaffinity beads with metal chelates as ligands and synthetic polymer beads as support matrices was prepared and tested. This study focused on developing metalโchelated poly(hydroxyethyl methacrylate) (PHEMA) gel beads in a spherical form (100โ150 ฮผm in diameter) f