𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Gel chromatography on a Sepharose 4B column: Earlier elution of protein-sodium dodecyl sulfate complexes of low Stokes radii

✍ Scribed by Pierre Wong; André Barbeau; Allen D. Roses


Publisher
Elsevier Science
Year
1985
Tongue
English
Weight
687 KB
Volume
146
Category
Article
ISSN
0003-2697

No coin nor oath required. For personal study only.

✦ Synopsis


A procedure is described for the determination of the Stokes radius of a detergent micelie by gel chromatography. It was observed that different lots of Sepharose -4B can exhibit a wide variation in the permeation of their gel pores. It is shown that this variation is due to differences in their pore size distribution. It has been observed that protein-sodium dodecyl sulfate (SDS) complexes of high Stokes radii eluted on a Sepharose 4B column with Stokes radii lower than the theoretical, as it has been previously reported but that protein-SDS complexes of low Stokes radii (~70 A), contrary to what might have been expected. eluted with Stokes radii higher than the theoretical. Evidence was obtained that their anomalous elution is due to an interaction of the detergent SDS with the gel pores of small diameter. cn 198j Academic PRESS. IX