๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Gas chromatographic determination of phenols as 2,4-dinitrophenyl ethers using glass capillary columns and an electron-capture detector

โœ Scribed by Matti Lehtonen


Publisher
Elsevier Science
Year
1980
Tongue
English
Weight
572 KB
Volume
202
Category
Article
ISSN
1873-3778

No coin nor oath required. For personal study only.

โœฆ Synopsis


An improved gas chromatographic method is described for the determination of all@-and alkoxy-substituted phenols as their 2,4-dinitiophenyl ethers. The gas chromatograpbic analysis is performed on QV-210 glass capillary or SP-2100 fused silica capillary c41Iumns using electron-capture detection. The separation of o-, mand p-substituted phenols and the isomers of dimethylphenols has been achieved. The sensitivity limit of electron-capture detection for various 2J-dinitrophenyl ethers is 0.01-0.09 ng injected and the coefficient of variation is about 5 %_

The reaction conditions for ether formation are described. The optimum pM, temperature and reaction time are 11.5, 50ยฐC and 40 min, respectively.


๐Ÿ“œ SIMILAR VOLUMES


An electron-capture gas chromatographic
โœ I.C. Cohen; B.B. Wheals ๐Ÿ“‚ Article ๐Ÿ“… 1969 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 757 KB

A method is described for the determination of residues of those substituted urea and carbamate herbicides which hydrolyze to give an aromatic amine. The 2,4clinitrophenyl derivatives of the amines, .which are amenable to gas chromatography and possess strong electron-captaring properties, are prepa