𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Fluorescence detection with an immersed flow cell in capillary electrophoresis

✍ Scribed by Kurosu, Yasuyuki ;Sasaki, Toru ;Saito, Muneo


Publisher
John Wiley and Sons
Year
1991
Tongue
English
Weight
407 KB
Volume
14
Category
Article
ISSN
0935-6304

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

An immersed flowcell has been developed for fluorescence detection in capillary electrophoresis (CE) by modifying an ordinary HPLC spectrofluorimetric detector flowcell. The minimum detectability obtained for riboflavine was about 4 fmol. The use of the flowcell to measure a fluorescence spectrum by the stopped‐flow method of shutting down the DC high voltage during a wavelength scan, has also been achieved successfully.