Flow system based on sequential delivery of air-sandwiched solutions into a micro cell for spectrophotometric catalytic analysis

✍ Scribed by Susumo Kawakubo; Masaaki Iwatsuki; Tsutomu Fukasawa


Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
647 KB
Volume
282
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

✦ Synopsis


An air-sandwiched sample, reagent solutions and water with a total volume of 317 PI are sequentially delivered by air carrier through PTFE tubing of 1 mm i.d. into a micro cell, where the solutions are mired to start reaction, and tbe absorbance is measured against reaction time. This flow system was compared with the stopped-flow/flowinjection and manual batch techniques with regard to the determination of 20-500 ng of sulphide by measuring the decolorization rate of Methyl Orange with bromate. In the proposed flow system, a micro sample is diluted only with limited small volumes of reagent solutions and free from the additional dilution due to dispersion of the reacting solution. Therefore, the absolute sensitivity was ten times that of the manual batch technique using a millilitre-level sample and three times that of the stopped-flow/flow-injection technique.