Fiber optic-based UV-absorption detector cell for high temperature open tubular column liquid chromatography

✍ Scribed by Lena M. Svensson; Karin E. Markides


Publisher
John Wiley and Sons
Year
1994
Tongue
English
Weight
462 KB
Volume
6
Category
Article
ISSN
1040-7685

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A new UV‐absorption detector cell for on‐column detection in open tubular column liquid chromatography was evaluated. Optical fibers and quick‐fit connections were employed for flexibility and ease in alignment and mounting in an oven for high temperature detection. Different optical fiber diameters were evaluated for 50 μm i.d. capillary columns. The best combination of optical fibers (200 μm source and collector) resulted in a detection limit of 2 × 10^−6^ M for acetophenone. No dramatic change in the signal‐to‐noise ratio was found when the temperature was raised to 150°C.