Feasibility of using real-time MRI to measure joint kinematics in 1.5T and open-bore 0.5T systems

✍ Scribed by Christine E. Draper; Juan M. Santos; Lampros C. Kourtis; Thor F. Besier; Michael Fredericson; Gary S. Beaupre; Garry E. Gold; Scott L. Delp


Publisher
John Wiley and Sons
Year
2008
Tongue
English
Weight
702 KB
Volume
28
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To test the feasibility and accuracy of measuring joint motion with real‐time MRI in a 1.5T scanner and in a 0.5T open‐bore scanner and to assess the dependence of measurement accuracy on movement speed.

Materials and Methods

We developed an MRI‐compatible motion phantom to evaluate the accuracy of tracking bone positions with real‐time MRI for varying movement speeds. The measurement error was determined by comparing phantom positions estimated from real‐time MRI to those measured using optical motion capture techniques. To assess the feasibility of measuring in vivo joint motion, we calculated 2D knee joint kinematics during knee extension in six subjects and compared them to previously reported measurements.

Results

Measurement accuracy decreased as the phantom's movement speed increased. The measurement accuracy was within 2 mm for velocities up to 217 mm/s in the 1.5T scanner and 38 mm/s in the 0.5T scanner. We measured knee joint kinematics with small intraobserver variation (variance of 0.8° for rotation and 3.6% of patellar width for translation).

Conclusion

Our results suggest that real‐time MRI can be used to measure joint kinematics when 2 mm accuracy is sufficient. They can also be used to prescribe the speed of joint motion necessary to achieve certain measurement accuracy. J. Magn. Reson. Imaging 2008;28:158–166. © 2008 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Measurements of occupational exposure to
✍ Sophie F. Riches; David J. Collins; Geoffrey D. Charles-Edwards; Jonathan C. Sha 📂 Article 📅 2007 🏛 John Wiley and Sons 🌐 English ⚖ 474 KB

## Abstract ## Purpose To determine if magnetic field exposure close to two clinical 1.5T magnetic resonance imaging (MRI) scanners during image acquisition and when moving in the spatially‐varying static magnetic field is compliant with European Union (EU) Directive 2004/40/EC (the Directive). #

Using real-time MRI to quantify altered
✍ Christine E. Draper; Thor F. Besier; Juan M. Santos; Fabio Jennings; Michael Fre 📂 Article 📅 2009 🏛 Elsevier Science 🌐 English ⚖ 218 KB

## Abstract Abnormal patellofemoral joint motion is a possible cause of patellofemoral pain, and patellar braces are thought to alleviate pain by restoring normal joint kinematics. We evaluated whether females with patellofemoral pain exhibit abnormal patellofemoral joint kinematics during dynamic,