𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Fast, pseudo-continuous arterial spin labeling for functional imaging using a two-coil system

✍ Scribed by Luis Hernandez-Garcia; Gregory R. Lee; Alberto L. Vazquez; Douglas C. Noll


Publisher
John Wiley and Sons
Year
2004
Tongue
English
Weight
466 KB
Volume
51
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A fast, two‐coil, pseudo‐continuous labeling scheme is presented. This new scheme permits the collection of a multislice subtraction pair in <3 s, depending on the subject's arterial transit times. The method consists of acquiring both control and tag images immediately after a labeling period that matches the arterial transit time. The theoretical basis of the technique, and simulations of the signal during changes in both transit time and perfusion are presented. Experimental data from functional imaging experiments were collected to demonstrate the technique and its characteristics. Magn Reson Med 51:577–585, 2004. © 2004 Wiley‐Liss, Inc.