Factors affecting dose distributions in brachytherapy of coronary arteries with P-32 linear source

✍ Scribed by Stanisław Pszona; Wojciech Bulski; Cezary Kępka; Adam Witkowski; Krzysztof Wincel; Barbara Zaręba


Publisher
Wielkopolskie Centrum Onkologii
Year
2006
Tongue
Polish
Weight
413 KB
Volume
11
Category
Article
ISSN
1507-1367

No coin nor oath required. For personal study only.

✦ Synopsis


Background

Long-term results of brachytherapy of coronary arteries may be infl uenced by inhomogeneity of the dose distribution. The dose distribution perturbations due to the presence of stent struts, guide wires and calcium deposits were therefore investigated.

Aim

The aim of the study was to investigate dose distribution perturbations due to the presence of stent struts, guide wire and calcium deposits during the irradiation. Both experimental and calculational techniques were applied to elucidate these infl uences.

Materials/Methods

The measurements were carried out with radiochromic foil in a phantom in the presence and absence of the stent and the guide wire. Due to the measurements being limited by the thickness of the detecting foil, the radial dose distribution with or without the metallic stent components were also performed using the MCNP Monte Carlo code. The same code was used to measure the infl uence of cholesterol plaques on dose distribution in the vessel walls.

Results

A large dose reduction in the area shielded by the guide wire was observed. The presence of the stent caused a 15% dose reduction behind the struts. The dose reduction caused by calcium deposits ranged from 11% to 32%.

Conclusions

These facts may increase the probability of late restenosis after irradiation due to underdosage. The inhomogeneity occurring behind the guide wire, stent struts and calcium deposits should be taken into account in the analysis of dose distributions for clinical applications.